Pseudomonophasic pulses with a gap need less current than symmetric pulses

Evidence

The difference in level between SYM-C and the various asymmetric pulse shapes RPS-C, RPS-IPG-C and TRI-C when they all produced the same response amplitude was determined. The response amplitude at which the pulse shapes were compared was equal to 50 % of the maximum amplitude of SYM-C. Given there was no effect of duration of deafness on the magnitude of these effects, the data were averaged across sessions for each animal and analyzed in a one-way repeated measures ANOVA. The star symbols in Fig. 4d show the mean across animal subjects. The effect of pulse shape was highly significant (F(3,18) = 22.76, p < 0.001). RPS-IPG required 2.2 dB less current than SYM (p < 0.05) while TRI required 2.7 dB more. There was no significant difference between SYM and RPS. — Dijk, Pim van, Physiology, Psychoacoustics and…, 2272