Deactivating pitch-indiscriminable cochlear implant electrodes yields no speech benefit

Evidence

Results indicated that there was no measurable benefit for deactivating channels based on pitch ranking for ACE users. Effects were negligible regardless of whether the deactivated channels were indiscriminable or discriminable based on pitch ranking. Care was taken to balance the design and ensure that all maps were used for sufficient time to obtain benefit. Prior to experimental map adjustments, clinical maps were optimised and each participant given an acclimatisation period prior to starting the study. There was a range in the number of channels selected for deactivation (1–6; median = four channels), which, when combined with the electrodes already switched off, resulted in a de-activation range of 1–10 (median = six channels). For an n-of-m strategy this number of deactivated electrodes may have been too small to demonstrate differences in performance; the selection of channels to deactivate was based purely on electrodes that were indiscriminable from one another in pitch ranking and for many users the number of indiscriminable electrode pairs was low so only small numbers were deactivated. The only statistically significant finding indicated that spectral resolution was poorer when electrodes were de-activated. When the channels were de-activated the rate of stimulation never changed, however the default filter allocation was used, which may have affected the spectral information (Henshall and McKay 2001 ). β€” Dijk, Pim van, Physiology, Psychoacoustics and…, 2110